Lumiscope 1143 manual

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28

Go to page of

A good user manual

The rules should oblige the seller to give the purchaser an operating instrucion of Lumiscope 1143, along with an item. The lack of an instruction or false information given to customer shall constitute grounds to apply for a complaint because of nonconformity of goods with the contract. In accordance with the law, a customer can receive an instruction in non-paper form; lately graphic and electronic forms of the manuals, as well as instructional videos have been majorly used. A necessary precondition for this is the unmistakable, legible character of an instruction.

What is an instruction?

The term originates from the Latin word „instructio”, which means organizing. Therefore, in an instruction of Lumiscope 1143 one could find a process description. An instruction's purpose is to teach, to ease the start-up and an item's use or performance of certain activities. An instruction is a compilation of information about an item/a service, it is a clue.

Unfortunately, only a few customers devote their time to read an instruction of Lumiscope 1143. A good user manual introduces us to a number of additional functionalities of the purchased item, and also helps us to avoid the formation of most of the defects.

What should a perfect user manual contain?

First and foremost, an user manual of Lumiscope 1143 should contain:
- informations concerning technical data of Lumiscope 1143
- name of the manufacturer and a year of construction of the Lumiscope 1143 item
- rules of operation, control and maintenance of the Lumiscope 1143 item
- safety signs and mark certificates which confirm compatibility with appropriate standards

Why don't we read the manuals?

Usually it results from the lack of time and certainty about functionalities of purchased items. Unfortunately, networking and start-up of Lumiscope 1143 alone are not enough. An instruction contains a number of clues concerning respective functionalities, safety rules, maintenance methods (what means should be used), eventual defects of Lumiscope 1143, and methods of problem resolution. Eventually, when one still can't find the answer to his problems, he will be directed to the Lumiscope service. Lately animated manuals and instructional videos are quite popular among customers. These kinds of user manuals are effective; they assure that a customer will familiarize himself with the whole material, and won't skip complicated, technical information of Lumiscope 1143.

Why one should read the manuals?

It is mostly in the manuals where we will find the details concerning construction and possibility of the Lumiscope 1143 item, and its use of respective accessory, as well as information concerning all the functions and facilities.

After a successful purchase of an item one should find a moment and get to know with every part of an instruction. Currently the manuals are carefully prearranged and translated, so they could be fully understood by its users. The manuals will serve as an informational aid.

Table of contents for the manual

  • Page 1

    WR IS T TY PE DI GI T AL B L OO D PR ES SU RE M ON IT OR MO DE L 11 43 IN ST R UC TI ON M AN U A L 1143-INS-LAB-RevB09 Important! Read and understand these instructions before using the Lumiscope Blood Pressure Monitor . If you do not understand any part of these instructions, contact your medical professional or GF dealer for direction in the use [...]

  • Page 2

    2 CONTENTS P ARTS AND COMPONENTS ............................................................................................................ 3 SYMBOLS ......................................................................................................................................... 4 1 GENERAL ................................................[...]

  • Page 3

    3 P ARTS AND COMPONENTS 1. Main Body 2. Display 3. Wrist Cuf f 4. Button ‘O/I’ 5. Button ‘M’ 6. Batteries 7. Storage Case[...]

  • Page 4

    4 SYMBOLS Symbol Meaning Symbol for the marking of electrical and electronics devices according to Directive 2002/96/EC. The device, accessories and packing must be disposed of as waste correctly at the end of the usage. Please follow Local Ordinances or Regulations for disposal. CE marking in conformity with EC directive 93/42/EEC Attention, consu[...]

  • Page 5

    5 1 GENERAL This instruction manual is intended to assist the user in safe and efficient operation of the wrist type digital blood pressure monitor (hereinafter: device) model 1 143. The device must be used in accordance with the procedures described in the manual. It is important to read and understand the entire manual, especially the section <[...]

  • Page 6

    6 Note: This device cannot provide reasonable accuracy if used or stored at temperature or humidity beyond the range stated in the <SPECIFICA TIONS> section of this manual.[...]

  • Page 7

    7 2 TIPS ON T AKING BLOOD PRESSURE MEASUREMENT 1. It is necessary to know that arterial blood pressure is subjected to sharp fluctuations. The level of the arterial blood pressure depends on many factors. Generally arterial blood pressure is lower in the summer and higher in the winter . Arterial blood pressure changes with atmosphere pressure and [...]

  • Page 8

    8 The illustration above is from The British Hypertension Society 2. Patients with cardiovascular diseases: T ake measurement under doctor’s instruction. Under no circumstances should you alter the dosage of any drug pre- scribed by your doctor! 3. Accurate measurement of blood pressure may be difficult in people with serious arteriosclerosis, we[...]

  • Page 9

    9 4. Keep quiet and relax during measurement. T o obtain accurate readings, wait one hour after a meal or smoking before taking a measurement. 5. W ait at least 5 minutes before taking the next measurement. Atherosclerosis and diabetes patients require longer intervals (10-15 minutes), since vessel elasticity may have decreased in these pathologies[...]

  • Page 10

    10 3 BA TTERY INST ALLA TION 1. Open the battery cover and install two alkaline AAA batteries in the battery compartment as indicated. Check polarity and ensure the batteries are correctly positioned; 2. Close the battery compartment cover . • Replacethe batteries whenthe replacement indicator appears in the display or when t[...]

  • Page 11

    1 1 4 CORRECT POSTURE FOR MEASUREMENT Sit down and place your elbow on a table. Ensure the wrist is at approximately the same level as the heart, and the forearm is extended naturally , palm up. Wrap the cuff around the left wrist as shown in the illustration, and ensure the cuff is attached to the wrist securely with the display facing the user . [...]

  • Page 12

    12 5 SETTING THE DA TE AND TIME The device has a 12-hour time format. The icon ‘AM’ indicates the time of Ante Meridiem and the icon ‘PM’ indicates the time of Post Meridiem. T o get accurate date and time, the user should preset the date and time correctly before the first use of this device. The operation procedure for presetting Date/T i[...]

  • Page 13

    13 3. Press button ‘M’ again to increase the number , and press button ‘O/I’ for confirmation; 4. When the year setup is confirmed, the month number will flash automatically . Follow the same instruction as above to set month, day and time; Fig.3 5. Press button ‘O/I’ to finish setup. If you want to change the date and time, repeat step[...]

  • Page 14

    14 6 T AKING A MEASUREMENT 1. Before taking a measurement, take 3 ~ 5 deep breaths and relax. Do not talk or move your arm during measurement; 2. Press button ‘O/I’, all symbols will display for 2 seconds as in Fig.4; then ‘0’ will display . The pump will then begin to inflate while displaying pressure. Generally the pressure will reach 190[...]

  • Page 15

    15 5. Press button ‘O/I’ to return to standby mode. W ait for at least 5 minutes before taking another measurement. If the device remains unused for 3 minutes, it will return to standby mode automatically . AUTOMA TIC INFLA TION This device has 4 levels of inflation pressure: 190 mmHg, 230 mmHg, 270 mmHg and 300 mmHg. When 190 mmHg is insuffici[...]

  • Page 16

    16 7 MEMOR Y FUNCTION MEMOR Y RECALL 1. Model 1 143 can store 90 sets of readings and will automatically calculate the average of the last 3 readings. When memory is full (90 sets of readings are stored), the oldest reading will be replaced by the newest. Memory will not clear even if the batteries are removed; 2. After a measurement is finished, o[...]

  • Page 17

    17 3. Press button ‘M’ again, the device displays ‘01’; then after 1-2 seconds the latest reading will display as in Fig.9; Fig.9 4. Press button ‘M’ again, the device displays ‘02’; then after 1-2 seconds the second-to-the-last reading will display . DELETING THE MEMORY T o clear memory: After a measurement is finished or when the [...]

  • Page 18

    18 8 WHO BLOOD PRESSURE CLASSIFICA TION The World Health Organization (WHO) has established standards for assessment of high or low blood pressure, regardless of age, as shown in the chart below: This device has an indicator located on the display which displays a segment, based on the current data, corresponding to the WHO classification. For exam[...]

  • Page 19

    19 Note: If the systolic blood pressur e and diastolic blood pressure fall into differ ent categories, use the higher value for classification.[...]

  • Page 20

    20 9 ERROR AND LOW BA TTERY INFORMA TION INDICA TION POSSIBLE REASON CORRECTION METHODS The wrist cuff is positioned incorrectly Movement of arm (or hand) or talking during measurement The wrist cuff is not inflated to sufficient pressure Arrhythmia Ensure wrist cuff is correctly positioned; repeat measurement Repeat measurement, following re co mm[...]

  • Page 21

    21 10 CARE, STORAGE, REP AIR AND RECYCLING 1. It is necessary to protect this device against high moisture, direct sunlight, shock, alcohol and gasoline. 2. Remove the batteries if the device is to be stored for a long period of time, and keep the batteries away from children. 3. Prevent sharp objects from coming into contact with the wrist cuff, a[...]

  • Page 22

    22 1 1 TROUBLESHOOTING SYMPTOM CHECK POINT REMEDY No display when button ‘O/I’ is pressed The batteries need to be replaced The batteries are incorrectly positioned The battery contacts are corroded Replace both batteries with new ones Install the batteries in correct position Clean the battery terminals with a dry cloth Inflation stops and re-[...]

  • Page 23

    23 T roubleshooting continued The reading is extremely low or high Is the wrist cuff at the same level as the heart? Is the wrist cuff wrapped correctly? Did you strain your arm (or hand) during measurement? Did you talk or move your arm (or hand) during measurement? Ensure your posture is correct Wrap the wrist cuff correctly Relax during measurem[...]

  • Page 24

    24 12 SPECIFICA TIONS Model 1 143 Size (L x W x H) 2 3 / 4 " x 2 3 / 8 " x 1 1 / 4 " (70 mm x 60 mm x 30 mm) Weight (without batteries) ~ .24 lb (1 10 g) Measuring method Oscillometry Measuring range 40 to 260 mmHg (blood pressure) 40 to 160 beats/minute (pulse rate) Measuring accuracy ± 3 mmHg for systolic and diastolic pressure ±[...]

  • Page 25

    25 Specifications continued Memory 90 sets of measurements T emperature and humidity Operation:50°F to104°F (10℃to 40℃), RH85% andbelow Storage:-4°F to122°F (-20℃to 50℃), RH85% andbelow Cuff size Fits wrist size 5" - 8" (127 mm - 203 mm ) Complete kit Main body , wris[...]

  • Page 26

    26 13 TWO-YEAR LIMITED W ARRANTY GF Health Products, Inc. guarantees this product free from defects in material and workmanship for a period of two years from the date of purchase, except as noted below: This product warranty does not cover damage caused by misuse or abuse, the attachment of any unauthorized accessory , alteration to the product, o[...]

  • Page 27

    27 W arranty is solely through the Repair Department. Service of this product by anyone other than the Repair Department voids warranty . Graham-Field ® and Lumiscope ® are registered trademarks of GF Health Products, Inc. Packaging, warranties, products, and specifications are subject to change without notice. GF Health Products, Inc. is not res[...]

  • Page 28

    GF Health Products, Inc. 2935 Nor theast Parkway Atlanta, Georgia 30360 telephone: 800-347-5678, 770-447-1609 fax: 800-726-0601, 678-291-3232 © October 2009 GF Health Products, Inc. Made in China www .grahamfield.com[...]