Milwaukee 2216-20 manual

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20

Go to page of

A good user manual

The rules should oblige the seller to give the purchaser an operating instrucion of Milwaukee 2216-20, along with an item. The lack of an instruction or false information given to customer shall constitute grounds to apply for a complaint because of nonconformity of goods with the contract. In accordance with the law, a customer can receive an instruction in non-paper form; lately graphic and electronic forms of the manuals, as well as instructional videos have been majorly used. A necessary precondition for this is the unmistakable, legible character of an instruction.

What is an instruction?

The term originates from the Latin word „instructio”, which means organizing. Therefore, in an instruction of Milwaukee 2216-20 one could find a process description. An instruction's purpose is to teach, to ease the start-up and an item's use or performance of certain activities. An instruction is a compilation of information about an item/a service, it is a clue.

Unfortunately, only a few customers devote their time to read an instruction of Milwaukee 2216-20. A good user manual introduces us to a number of additional functionalities of the purchased item, and also helps us to avoid the formation of most of the defects.

What should a perfect user manual contain?

First and foremost, an user manual of Milwaukee 2216-20 should contain:
- informations concerning technical data of Milwaukee 2216-20
- name of the manufacturer and a year of construction of the Milwaukee 2216-20 item
- rules of operation, control and maintenance of the Milwaukee 2216-20 item
- safety signs and mark certificates which confirm compatibility with appropriate standards

Why don't we read the manuals?

Usually it results from the lack of time and certainty about functionalities of purchased items. Unfortunately, networking and start-up of Milwaukee 2216-20 alone are not enough. An instruction contains a number of clues concerning respective functionalities, safety rules, maintenance methods (what means should be used), eventual defects of Milwaukee 2216-20, and methods of problem resolution. Eventually, when one still can't find the answer to his problems, he will be directed to the Milwaukee service. Lately animated manuals and instructional videos are quite popular among customers. These kinds of user manuals are effective; they assure that a customer will familiarize himself with the whole material, and won't skip complicated, technical information of Milwaukee 2216-20.

Why one should read the manuals?

It is mostly in the manuals where we will find the details concerning construction and possibility of the Milwaukee 2216-20 item, and its use of respective accessory, as well as information concerning all the functions and facilities.

After a successful purchase of an item one should find a moment and get to know with every part of an instruction. Currently the manuals are carefully prearranged and translated, so they could be fully understood by its users. The manuals will serve as an informational aid.

Table of contents for the manual

  • Page 1

    MIL W AUKEE ELECTRIC T OOL CORPORA TION – TEST & MEASUREMEN T PRODUCT REP AIR SER VICE INSTRUCTIONS — CALIBRA TION 2 216- 20 Light Commercial Digital Multimeter Environmental Condition Perform all calibration at an ambient temperature of 23°C ± 2°C / 73.4°F ± 3.6°F a nd relative humidity of 80% - allow the digital multimeter to si t a[...]

  • Page 2

    i T ABLE OF CONTENTS Page Introduction 1 Precautions and Safety Information 1 Symbols 1 Safety 2 S pecifications 3 General S pecification 3 V oltage S pecif ication 4 Current S pecifications 5 Resist ance, Continuity , Capacit ance S pecifications 6 Frequency , Rotary Switch, Auto Power Of f S pecification 7 Physical and environment characteristics[...]

  • Page 3

    1 Introduction W arning T o avoid shock or injury , do not perform the verification test s or calibration procedu res described in the manual unless y ou are qualified to do so. The information provided in this document is for the use of qu alified personnel only . Caution The 2216-20 cont ain p art s that can be damaged by st atic discharge. Follo[...]

  • Page 4

    2 SAFETY Review the following safety precaution s to avoid injury and prevent damag e to this product or any product s connected to it. T o avoid potential hazards, use the product only as spe cified. CAUTION: These st atement s identify conditions or practices that could result in damage to the equipment or other pro perty . W ARNING: These statem[...]

  • Page 5

    3 SPECIFICA TIONS All specifications are warranted unless noted typical and apply to the 2216 -20 DMM. S tated accuracies are at 23°C±5°C at tha n 80% relative humidity and without the battery indicator displayed. Gen era l sp ec ifi ca tions n o i t p i r c s e D s c i t s i r e t c a r a h C s t n u o c 0 0 0 6 t n u o c y a l p s i D c e s / [...]

  • Page 6

    4 Measurement Characteris tics Accuracy is ±(% reading + number of digit s) at 23°C ± 5°C, less than 80% R.H. T emperature coeffici ent: 0.1 * (S pecifi ed accuracy)/°C , < 18°C , > 28°C (1) DC V olt age Range Resolution Accuracy 600.0mV 0.1mV 6.000V 0.001V 60.00V 0.01V 600.0V 0.1V ± (0.5% reading + 2 digits) Input Impedance: 10M Ω [...]

  • Page 7

    5 (3) DC Current Range Resolution Accuracy 6.000A 0.001A 10.00A 0.01A ± (1% reading + 3 digit s) Response Ti me: < 1.5 sec. Maximum measurement time: 5 minutes at 10A with at least 20 minutes rest time. Overload Protection: AC/D C 20A for 10 sec ond. (4) AC Current Range Resolution Accuracy (Sine W av e) 6.000A * 0.001A 10.00A 0.01A ± (1.5% re[...]

  • Page 8

    6 (6) Continuity Range Resolution Accuracy 600.0 Ω 0.1 Ω ± (1% reading + 5 digit s) Open Circuit V olt age: Approx. 0.6V Continuity: Built-in buzzer sounds whe n measured resistance i s less than 30 Ω and sounds off when measured resist ance is more than 200 Ω , Between 30 Ω to 200 Ω the buzzer maybe sound or of f either . Continuity I[...]

  • Page 9

    7 (1 1) Auto Po wer Off The meter au tomatically turns the power off after powering up and no operation for 20 ± 5 minutes. T o power up, turn the knob back to OFF f irst. (12) Measuring Rate: 3 t i mes / sec. (13) Fuse: 10A/600V Fast Action Fuse (1 4) Batt er y : Alkaline 1.5V Battery * 2 (15) Low Battery V oltage: 2.3V ± 0.1V (16) Backlight The[...]

  • Page 10

    8 Physical and Environment al Characteri stics n o i t p i r c s e D s c i t s i r e t c a r a h C Dime nsio ns (H ×W ×D) 170mm × 80mm × 57m m g K 3 . 0 ) y r e t t a b h t i w ( t h g i e W n o i t p i r c s e D s c i t s i r e t c a r a h c l a t n e m n o r i v n E C ° 0 5 + o t 0 1 - g n i t a r e p o e r u t a r e p m e T Non-Operating -4[...]

  • Page 11

    9 Required Equipment Required equipment is listed in T able B. If the recommended models are not available, equip ment with equivalent specifications may be used. Repairs or se rvicing should be per formed only by qualified personn el. T able B. Required Equipment Equipment Required Characteristics Recommended Model Calibrator AC Voltage Range: 0 ?[...]

  • Page 12

    10 Basic Maintenance W arning T o avoid shock, remove the test leads an d any input signals before opening the case o r replacing the battery or fuses. Opening the Meter Ca se Caution T o avoid unintentional shock circuit, always place the uncovered Meter assembl y on a protective surface. When the ca se of the Meter is open, circuit connections ar[...]

  • Page 13

    11 T esting Fuses T o test the internal fuses of the meter . 1. T urn the rotary selector s witch to the Ω position. 2. T o test FS1, plug a test lead into V Ω Hz input terminal, and touch the probe to the A input terminal. The displ ay should indicate between 0.0 to 0.2 Ω . (Bussmann KTK-10 recommended). If display reads higher than 0.2 Ω [...]

  • Page 14

    12 Performance T est s The following performance test s verify the co mplete operability of the Meter an d check the accuracy of each Meter function against the Meter’s specifications. Accuracy specificatio ns are valid for a period of one year after calibration, when measured at an operating temperature of 18°C to 28°C and at A maximum of 80% [...]

  • Page 15

    13 T esting the V olt age Function To verify accuracy in the AC and DC voltage ranges, do the following: 1. T urn t he ro t ar y switch to “ V ” position . 2. Connect the Calibrator to the V Ω and CO M inputs on the Meter. 3. Set the Calibrator for the voltage and frequency from step 1 to 12 in Table 1. 4. Compare the reading on the Meter dis[...]

  • Page 16

    14 T esting the Resist ance Function T o verify the accuracy of the resist ance function, do the following: 1. Connect the Calibrator to V Ω Hz and CO M on t he Me ter . 2. T u rn t he ro tary swi tch to Ω . 3. Apply the inputs for st ep 1-6 in T able 3. 4. Compare the Meter display readings to the displ ay readings in T able 3. 5. If the displa[...]

  • Page 17

    15 T esting the Ampere Function T o verify the accuracy of AC current measurement function s, do the following: 1. Connect the Calibrator to the A and CO M inpu t s on t he Me ter . 2. T u rn the ro ta ry swit ch to A~. 3. Apply the inputs for step s 1-4 in T able 5. 4. For each input, comp are the readings on the Meter display to the reading in T [...]

  • Page 18

    16 T able 7 Frequency T est: Step Source Level Waveform Readin g 1 8.0Vp-p 2.00Hz Square Wa ve 1.98Hz to 2.02Hz 2 8.0Vp-p 80.00Hz Square Wa ve 79.90Hz to 80.10Hz 3 8.0Vp-p 800.0Hz Square Wa ve 799.0Hz to 801.0Hz 4 8.0Vp-p 8.000kHz Square Wa ve 7.990kHz to 8.010kHz 5 10.0Vp-p 50.00kHz Square Wave 49.93kHz to 50.07kHz 6 230.0Vp-p 10.00 Hz Sine Wave 9[...]

  • Page 19

    17 Calibration Procedure Recalibrate the meter: It is recommended that the meter may be calibrated once year . Use the following procedure to calibrate the meter . Envir on ment a l C ondi t ion 1. Perform calibration at an ambient temperature and relative humidity (23°C±2°C a nd R.H. 80%). Allow the instrument to sit at this te mperature for at[...]

  • Page 20

    18 (E) DCA Calibration (V R7) 1. Rotate the rotary switch to DCA Function. 2. Connect the output of a DCA calibrator to V and COM input terminals of the meter. 3. Output DCA 5.000A from the calibrator. 4. Adjust VR7 until the display shows 5.000A ± 0.001A. 5. Turn off the output of the calibrator an d disconnect the meter from the calib rator. (F)[...]