Leica RM2125 manual

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15

Go to page of

A good user manual

The rules should oblige the seller to give the purchaser an operating instrucion of Leica RM2125, along with an item. The lack of an instruction or false information given to customer shall constitute grounds to apply for a complaint because of nonconformity of goods with the contract. In accordance with the law, a customer can receive an instruction in non-paper form; lately graphic and electronic forms of the manuals, as well as instructional videos have been majorly used. A necessary precondition for this is the unmistakable, legible character of an instruction.

What is an instruction?

The term originates from the Latin word „instructio”, which means organizing. Therefore, in an instruction of Leica RM2125 one could find a process description. An instruction's purpose is to teach, to ease the start-up and an item's use or performance of certain activities. An instruction is a compilation of information about an item/a service, it is a clue.

Unfortunately, only a few customers devote their time to read an instruction of Leica RM2125. A good user manual introduces us to a number of additional functionalities of the purchased item, and also helps us to avoid the formation of most of the defects.

What should a perfect user manual contain?

First and foremost, an user manual of Leica RM2125 should contain:
- informations concerning technical data of Leica RM2125
- name of the manufacturer and a year of construction of the Leica RM2125 item
- rules of operation, control and maintenance of the Leica RM2125 item
- safety signs and mark certificates which confirm compatibility with appropriate standards

Why don't we read the manuals?

Usually it results from the lack of time and certainty about functionalities of purchased items. Unfortunately, networking and start-up of Leica RM2125 alone are not enough. An instruction contains a number of clues concerning respective functionalities, safety rules, maintenance methods (what means should be used), eventual defects of Leica RM2125, and methods of problem resolution. Eventually, when one still can't find the answer to his problems, he will be directed to the Leica service. Lately animated manuals and instructional videos are quite popular among customers. These kinds of user manuals are effective; they assure that a customer will familiarize himself with the whole material, and won't skip complicated, technical information of Leica RM2125.

Why one should read the manuals?

It is mostly in the manuals where we will find the details concerning construction and possibility of the Leica RM2125 item, and its use of respective accessory, as well as information concerning all the functions and facilities.

After a successful purchase of an item one should find a moment and get to know with every part of an instruction. Currently the manuals are carefully prearranged and translated, so they could be fully understood by its users. The manuals will serve as an informational aid.

Table of contents for the manual

  • Page 1

    The Leica RM2125 Microtome Online T roubleshooting Guide[...]

  • Page 2

    Leica RM2125 T roubleshooting Guide T able of Contents Problem Page 1. Thick and thin sections 1-2 2. Compressed sections (sections are extremely compressed, wrinkled or jammed together) 3-4 3. Chatter 5-6 4. Scratches or splits in sections 7 5. No specimen advance (no section produced) 8 6. Drifting of handwheel 9 7. Specimen is picked up in the r[...]

  • Page 3

    Leica RM2125 T roubleshooting Guide Problem 1. Thick and thin sections a) Partial sections b) Shorter sections in ribbon c) Section thickness varies from one section to another d) Skipping sections e) V aried thickness in one individual section Cause(s) / Corrective Action I) The clearance angle is too small. a) Systematically try several clearance[...]

  • Page 4

    Leica RM2125 T roubleshooting Guide Problem 1. Thick and thin sections (continued) a) Partial sections b) Shorter sections in ribbon c) Section thickness varies from one section to another d) Skipping sections e) V aried thickness in one individual section Cause(s) / Corrective Action VI) Backlash in specimen advance system. l) T est for backlash b[...]

  • Page 5

    Leica RM2125 T roubleshooting Guide Problem 2. Compressed sections (sections are extremely compressed, wrinkled or jammed together) a) Compressed b) Wrinkled c) Jammed Cause(s) / Corrective Action I) Dull knife/blade. a) Use a different part of the cutting edge or use a new knife/blade. II) Blade/knife is gummed up with paraffin b) Clean knife with[...]

  • Page 6

    Leica RM2125 T roubleshooting Guide Problem 2. Compressed sections (sections are extremely compressed, wrinkled or jammed together) (continued) a) Compressed b) Wrinkled c) Jammed Cause(s) / Corrective Action VII) Knife blade is not parallel with pressure plate. (cont.) l) Adjust Pressure Plate with set screws at the bottom of the knife holder (tur[...]

  • Page 7

    Leica RM2125 T roubleshooting Guide Problem 3. Chatter a) V enetian Blind affect b) W ash board c) Undulations Cause(s) / Corrective Action I) Sectioning speed too high. a) Rotate the handwheel at a slower speed. II) Clearance angle too big. c) Clearance angle adjustment: Systematically decrease the clearance angle until the optimum adjustment is o[...]

  • Page 8

    Leica RM2125 T roubleshooting Guide Problem 3. Chatter (continued) a) V enetian Blind affect b) W ash board c) Undulations Cause(s) / Corrective Action – Decalcification – Paraffin is too old – Paraffin impurities IX) T abletop stability . k) Move the microtome or table to a more stable location, or replace the table if the table is flimsy . [...]

  • Page 9

    Leica RM2125 T roubleshooting Guide Problem 4. Scratches or splits in sections Cause(s) / Corrective Action I) Knife: – Defect in knife edge – Hard specimen in block – Paraffin on front or back of knife edge – Possible staple left in paraffin a) Move to a new area of the knife and observe if the scratches move with it. If so, replace the kn[...]

  • Page 10

    Leica RM2125 T roubleshooting Guide Problem 5. No specimen advance (no section produced) Cause(s) / Corrective Action I) Specimen reached the front feed limit. a) T urn the coarse feed wheel in the appropriate direction to move the specimen towards the rear limit. II) User may accidently be holding the coarse feed wheel while sectioning. III) The c[...]

  • Page 11

    Leica RM2125 T roubleshooting Guide Problem 6. Drifting of handwheel Cause(s) / Corrective Action I) Hand wheel is not balanced properly . Handwheel should remain at 12 o’clock when positioned at that location. No drifting should occur . a) Adjust the handwheel position and/or collar sleeve position. Service call is needed . 9 back to T able of C[...]

  • Page 12

    Leica RM2125 T roubleshooting Guide Problem 7. Specimen is picked up in the return stroke of the specimen arm Cause(s) / Corrective Action I) Static electricity charge may be built up on the knife holder or specimen head. a) Adjust the temperature or humidity level in the room. b) Use an ionizer to neutralize the static charge. c) Clean components [...]

  • Page 13

    Leica RM2125 T roubleshooting Guide Problem 8. Sections sticking to pressure plate Cause(s) / Corrective Action I) T oo much paraffin. a) Clean the knife with a swab moistened with Xylene. Stroke the swab in an upward motion away from the blade edge, not down on the blade edge. II) Dirty pressure plate. b) Clean the plate with a swab moistened with[...]

  • Page 14

    Leica RM2125 T roubleshooting Guide Problem 9. Unable to retract specimen (Note: This is only on the Leica RM2125 RT model) Cause(s) / Corrective Action I) Retraction mechanism on specimen advance system is defective. a) Schedule a service call. 12 back to T able of Contents[...]

  • Page 15

    Leica RM2125 T roubleshooting Guide Problem 10. Microtome makes "clunking noise" when sectioning Cause(s) / Corrective Action I) Linear bearings in cage are at their lower limits. a) Linear bearings need to be cleaned and possibly re-greased. II) Alignment of bearing cage is incorrect. b) Schedule a service call. 13 back to T able of Cont[...]